Display Shows Possible Cause(s) Possible Solution(s)
Invalid Hyper
Values
j The hyper warning limit value
must be greater than all of
your target ranges in the time
block settings.
j Reset the hyper warning limit or revise the target
ranges in time blocks and re-enter the hyper
warning limit.
j Enter a hyper warning limit that is above the target
ranges of your time blocks. See Chapter 5,
“Changing Meter Settings.
Invalid Hypo Values j The hypo warning limit value
must be less than all of your
target ranges in the time
block settings.
j Reset the hypo warning limit or revise the target
ranges in time blocks and re-enter the hypo
warning limit.
j Enter a hypo warning limit that is below the target
ranges of your time blocks. See Chapter 5,
“Changing Meter Settings.
Invalid PIN j The wrong PIN was entered. j Select OK and re-enter the PIN shown on the pump
display.
Invalid Record
Time/Date
j The time/date entered is
invalid (Add Data entries
cannot be set to occur in the
future).
j Re-enter time/date.
50813_00048009694_D.indb 240 7/3/12 3:13 PM