  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 242
241
8
Display Shows Possible Cause(s) Possible Solution(s)
Invalid Target
Range
j The lower target range value
is above the upper target
range value.
j Reset the target range values.
j Enter the correct lower target range value and
upper target range value. See Chapter 5, “Changing
Meter Settings.
j The range selected conflicts
with your hyper and/or hypo
warning limit settings.
j Reset range or revise warning limit settings and
re-enter range.
j Enter the correct lower target range value and
upper target range value. See Chapter 5, “Changing
Meter Settings.
LO bG Warning j Your blood glucose may be
lower than the measuring
range of the system.
j Treat your low blood glucose as recommended by
your healthcare professional.
j If you are experiencing any of the common
symptoms of low blood glucose, contact your
healthcare professional immediately.
Meter and Pump
Not Paired
j Meter and pump have not
been paired. You cannot use
the pump functions without
pairing the meter and the
pump.
j You must pair the meter and pump to use these
features.
j For instructions on how to pair the meter and pump,
see the Advanced Owner’s Booklet.
Meter Battery Low j Battery power is low. j Install new batteries. See Chapter 7, “Care and
Maintenance.